Product Details
5 kV Digital Insulation Tester by Fluke Electronics®. he Fluke 1550C performs digital insulation testing up to 5 kV, and makes data collection easy through the Fluke Connect™ wireles app. Remote operation lets you spend as little time as posible in potentially hazardous zones. Real-time trending shows a live graph on your phone, rather than a hard-to-read set of changing values. And you can create reports quickly, using pre-designed templates that you can customize with company logos. It’s the perfect tool for preventive or predictive maintenance programs designed to identify potential equipment failures before they occur. And it comes with a three-year warranty and CAT IV 600 V safety rating.
Specifications
Test voltage (DC): |
250 V: Range< 250 kΩ,250 kΩ to 5 GΩ, 5 GΩ to 50 GΩ> 50 GΩ |
Accuracy (± reading): Unspecified, 5%, 20%, Unspecified |
500 V: |
Range: < 500 kΩ, 500 kΩ to 10 GΩ, 10 GΩ to 100 GΩ, > 100 GΩ |
Accuracy (± reading): Unspecified, 5%, 20% |
1000 V: |
Range: < 1 MΩ, 1 MΩ to 20 GΩ, 20 GΩ to 200 GΩ, > 200 GΩ |
2500 V: |
Range: < 2.5 MΩ, 2.5 MΩ to 50 GΩ, 50 GΩ to 500 GΩ, > 500 GΩ |
5000 V: |
Range: < 5 MΩ, 5 MΩ to 100 GΩ, 100 GΩ to 1 TΩ, > 1 TΩ |
Bar graph range: 0 to 1 TΩ (1550C) - 0 to 2 TΩ (1555) |
Insulation test voltage accuracy: -0%, +10% at 1 mA load Current |
Induced AC mains Current rejection: 2 mA maximum |
Charging rate for capacitive load: 5 seconds per μF |
Discharge rate for capacitive load: 1.5 s/μF |
Leakage Current measurement: Range 1 nA to 2 mA |
Accuracy: ± (5% + 2 nA) |
Capacitance measurement: |
Range: 0.01 μF to 15.00 μF |
Accuracy: ±: (15% rdg + 0.03 μF) |
Timer: |
Range: 0 to 99 minutes |
Resolution setting: 1 minute |
Resolution indication: 1 second |
Live circuit warning |
Warning range: 30 V to 660 V AC/DC, 50/60 Hz |
Voltage accuracy: ± (15% + 2 V) |
Software specifications: FlukeView Forms™ basic software requires a PC running Windows XP, Windows Vista , Windows 7 or Windows 8 |
Display: 75 x 105 mm |
Power: 12 V lead-acid rechargeable battery, Yuasa NP2.8-12 |
Charger Input (AC): 85 V to 250 V AC 50/60 Hz 20 VA |
Dimensions: 269 x 277 x 160 mm (10.6 x 10.9 x 6.3 in) |
Weight: 3.7 kg (8.2 lb) |
Temperature (operating): -20°C to 50°C (-4°F to 122°F) |
Temperature (storage): -20°C to 65°C (-4°F to 149°F) |
Humidity: 80% to 31°C decreasing linearly to 50% at 50°C |
Altitude: 2000 m |
Enclosure sealing: IP40 |
Input overload protection: 1000 V AC |
Electromagnetic compatibility: EN 61326 |
Safety compliance: IEC 61010-1: Overvoltage category II, Pollution Degree 2, IEC 61010-2-030: Measurement 600V CAT IV / 1000V CAT III, IEC 61557-1, IEC 61557-2 |
Features
- Quickly configure, start and stop tests on the 1550C remotely, a safe distance away from any operating, energized equipment, using Fluke Connect® and a smart phone
- Real time trending gives you easy-to-see visual cues, rather than hard to read, rapidly changing values
- Report creation using Fluke Connect desktop saves time and makes test results easier to understand
- Fluke Connect saves test results to the cloud for analysis and preventive maintenance
Fluke Electronics® Corporation is the world leader in the manufacture, distribution and service of electronic test tools and software. Since its founding in 1948, Fluke has helped define and grow a unique technology market, providing testing and troubleshooting capabilities that have grown to mission critical status in manufacturing and service industries. Every new manufacturing plant, office, hospital, or facility built today represents another potential customer for Fluke products.